The effect of body mass index on complications in cardiac implantable electronic device surgery
Loading...
Date
2023
Journal Title
Journal ISSN
Volume Title
Publisher
WILEY
Open Access Color
Green Open Access
No
OpenAIRE Downloads
OpenAIRE Views
Publicly Funded
No
Abstract
Background: Cardiac implantable electronic device (CIED) procedures are prone to complications. In our study, we investigated the effect of body mass index (BMI) on CIED-related complications.
Methods: 1676 patients who had undergone CIED surgery (de novo implantation, system upgrade, generator change, pocket revision or lead replacement) at two heart centers in Turkey and met the study criteria were included in our study. For analysis of primary and secondary endpoints, patients were classified as non-obese (BMI < 25 kg/m2), overweight (25 ≤ BMI < 30 kg/m2), and obese (BMI ≥ 30 kg/m2). The primary endpoint was accepted as cumulative events, including the composite ofclinically significant hematoma (CSH), pericardial effusion or tamponade, pneumoth- orax, and infection related to the device system. Secondary outcomes included each component of cumulative events.
Results: The rate of cumulative events, defined as primary outcome, was higher in the obese patient group, and we found a significant difference between the groups (3.0%, 4.3%, 8.9%, p = .001). CSH and pneumothorax rates were significantly higher in the obese patient group (0.3%, 0.9%, 1.9%, p = .04; 1.0%, 1.4%, 3.3%, p = .04, respectively). According to our multivariate model analysis; gender (OR:1.882, 95%CI:1.156–3.064, p = .01), hypertension (OR:4.768, 95%CI:2.470–9.204, p < .001), BMI (OR:1.069, 95%CI:1.012–1.129, p = .01) were independent predictors of cumulative events rates. Conclusions: Periprocedural complications associated with CIED (especially hematoma and pneumothorax) are more common in the group with high BMI.
Description
ORCID
Keywords
Cardiovascular implantable electronic devices, Complication, Pacemaker, Body mass index, Hematoma, Pacemaker, Artificial, body mass index; cardiovascular implantable electronic devices; complication; pacemaker, Cardiovascular implantable electronic devices, Pneumothorax, Pericardial Effusion, Defibrillators, Implantable, Body Mass Index, Pacemaker, Humans, Obesity, Complication, Body mass index, Retrospective Studies, body mass index, complication, pacemaker, Risk-Factors, cardiovascular implantable electronic devices, Infection, Pocket Hematoma, Cardiovascular Implantable Electronic Devices
Turkish CoHE Thesis Center URL
Fields of Science
Citation
Güzel, T., Demir, M., Aktan, A., Kılıç, R., Arslan, B., Günlü, S., Altıntaş, B., Karahan, M. Z., Özbek, M., Aslan, B., Arpa, A., Coşkun, M. S., Altunbaş, M., Tüzün, R., Akgümüş, A., Karadeniz, M., Aydın, S., Güzel, H., Aslan, S. F., Söner, S., … Ertaş, F. (2023). The effect of body mass index on complications in cardiac implantable electronic device surgery. Pacing and clinical electrophysiology : PACE, 10.1111/pace.14903. Advance online publication. https://doi.org/10.1111/pace.14903
WoS Q
Q3
Scopus Q
Q3

OpenCitations Citation Count
1
Source
Pacing Clin Electrophysiol
Volume
47
Issue
Start Page
1
End Page
8
PlumX Metrics
Citations
Scopus : 2
Captures
Mendeley Readers : 6
SCOPUS™ Citations
2
checked on Feb 06, 2026
Web of Science™ Citations
2
checked on Feb 06, 2026
Page Views
18
checked on Feb 06, 2026
Downloads
52
checked on Feb 06, 2026
Google Scholar™